Clinical Radiology
Volume 64, Issue 10 , Pages 1009-1014, October 2009

Needle placement accuracy during stereotactic localization mammography

  • D.H. Green

      Affiliations

    • Corresponding Author InformationGuarantor and correspondent: D.H. Green, Breast Screening Unit, Elm House, Wirral University Teaching Hospital, Wirral, CH63 4JY, UK. Tel.: +44 151 482 7661; fax: +44 151 482 7619.

Breast Screening Unit, Wirral University Teaching Hospital, Wirral, UK

Received 8 October 2008; received in revised form 14 May 2009; accepted 2 June 2009.

Aim

To derive a mathematical model to describe the relationship between lesion position in the breast and measurements derived from the stereoradiographs to enable more accurate sampling of a lesion during stereotactic mammographic needle placement.

Materials and methods

The affect that registration errors have on the accuracy of needle placement when identifying the lesion on the stereoradiographs was investigated using the mathematical model.

Results

The focus-to-film distance of the x-ray tube and the horizontal distance of the lesion from the centre of rotation have little effect on error. Registration errors for lesions lying at a greater perpendicular distance in the breast from the centre of rotation produce smaller localization errors when compared with lesions sited closer. Lesion registration errors during marking of the stereoradiographs are exacerbated by decreasing the angle of x-ray tube swing.

Conclusions

When problems are encountered in making an accurate registration of the lesion on the stereoradiographs, consider the following error reducing strategies: (1) employ an approach that places the lesion the maximum distance away from the film cassette; (2) avoid reducing the angle of tube swing; and (3) consider sampling superficial and deep to, as well as at, the location indicated. The possibility of erroneous tissue sampling should be borne in mind when reviewing the pathology report.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0009-9260(09)00210-4

doi:10.1016/j.crad.2009.06.001

Clinical Radiology
Volume 64, Issue 10 , Pages 1009-1014, October 2009