Absorbed radiation dose of the female breast during diagnostic multidetector chest CT and dose reduction with a tungsten–antimony composite breast shield: preliminary results

Medical College of Virginia Hospitals-VCU Health System, Richmond, VA, USA

Received 5 April 2007; received in revised form 25 June 2007; accepted 1 July 2007.

Aim

To determine the absorbed radiation dose to the female breast during chest computed tomography (CT), and whether a custom-designed breast shield can reduce that dose.

Materials and methods

Bilateral breast phantoms were combined with an anthropomorphic torso phantom. Each breast phantom contained 20 thermoluminescent dosimeter (TLD) cavities. Eight cavities were used per phantom. Absorbed radiation was measured using TLD 100s. Three-stacked TLDs comprised a set. Three sets of three TLDs were positioned at eight designated locations and three depths (surface; 1cm; 4cm). One set of three TLDs was positioned at eight additional designations, 1cm deep. Each breast was divided anatomically into quadrants. In total, 32 TLD sets/96 TLDs were deployed. The breast–torso phantom was consecutively imaged using a 16-detector array CT machine. Subsequently, 32 new TLD sets were similarly placed, the phantom re-imaged in a likewise manner, but with the application of a tungsten–antimony composite breast shield. TLD readings were averaged and calculated.

Results

Average absorbed radiation doses for unshielded right and left breast phantoms ranged from 13.83–19.36mGy, and 14–20.47mGy, respectively. The absorbed dose in the shielded right and left breast was reduced to 6.64–8.12mGy, and 6.7–8.03mGy, respectively. Average absorbed radiation doses based on the depth for the unshielded breasts ranged from 15.4–18.3mGy. Shielding reduced this dose to 7–7.9mGy. Unshielded absorbed radiation doses based on anatomic quadrants ranged from 17.5–18.9mGy. Shielding reduced this dose to 7–7.5mGy.

Conclusions

The average absorbed radiation dose to the unshielded female breast phantom is approximately 14–20mGy. An externally applied shield can reduce this absorbed dose by 56–61%.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0009-9260(07)00430-8

doi:10.1016/j.crad.2007.07.029

Clinical Radiology
Volume 63, Issue 3 , Pages 278-288, March 2008